会议论文详细信息
7th International Conference on 3D Radiation Dosimetry
Comprehensive quality assurance for base of skull IMRT
Thomas, A.^1 ; O'Daniel, J.^1 ; Adamovics, J.^2 ; Ibbott, G.^3 ; Oldham, M.^1
Duke University Medical Center, Durham, NC, United States^1
Rider University, Lawerenceville, NJ, United States^2
University of Texas - MD Anderson, Houston, TX, United States^3
关键词: Color map;    Comprehensive qualities;    Dose distributions;    Head and neck;    High quality;    High-accuracy;    Optical CT;    Treatment plans;   
Others  :  https://iopscience.iop.org/article/10.1088/1742-6596/444/1/012050/pdf
DOI  :  10.1088/1742-6596/444/1/012050
来源: IOP
PDF
【 摘 要 】
Six base of skull IMRT treatment plans were delivered to Presage dosimeters within the RPC Head and Neck Phantom for quality assurance (QA) verification. Isotropic 2mm 3D data were acquired by optical-CT scanning with the DLOS system (Duke Large Optical-CT Scanner) and compared to the Eclipse (Varian) treatment plan. Normalized Dose Distribution (NDD) pass rates were obtained for a number of criteria. High quality 3D dosimetry data was observed from the DLOS system, illustrated here through colormaps, isodose lines, and profiles. Excellent agreement with the planned dose distributions was also observed with NDD analysis revealing > 90% pass rates (with criteria 3%, 2mm), and noise
【 预 览 】
附件列表
Files Size Format View
Comprehensive quality assurance for base of skull IMRT 894KB PDF download
  文献评价指标  
  下载次数:9次 浏览次数:37次