会议论文详细信息
| 7th International Conference on 3D Radiation Dosimetry | |
| 3D Dose reconstruction: Banding artefacts in cine mode EPID images during VMAT delivery | |
| Woodruff, H.C.^1 ; Greer, P.B.^1,2 | |
| School of Medicine and Public Health, University of Newcastle, Newcastle, NSW 2308, Australia^1 | |
| Calvary Mater Newcastle Hospital, Newcastle, NSW 2298, Australia^2 | |
| 关键词: Dose reconstruction; Field images; Pulse sequence; | |
| Others : https://iopscience.iop.org/article/10.1088/1742-6596/444/1/012042/pdf DOI : 10.1088/1742-6596/444/1/012042 |
|
| 来源: IOP | |
PDF
|
|
【 摘 要 】
Cine (continuous) mode images obtained during VMAT delivery are heavily degraded by banding artefacts. We have developed a method to reconstruct the pulse sequence (and hence dose deposited) from open field images. For clinical VMAT fields we have devised a frame averaging strategy that greatly improves image quality and dosimetric information for three-dimensional dose reconstruction.
【 预 览 】
| Files | Size | Format | View |
|---|---|---|---|
| 3D Dose reconstruction: Banding artefacts in cine mode EPID images during VMAT delivery | 720KB |
PDF