| 7th International Conference on 3D Radiation Dosimetry | |
| Basic investigations on LCV micelle gel | |
| Ebenezer, S.B.^1 ; Rafic, M.K.^1 ; Ravindran, P.B.^1 | |
| Department of Radiation Oncology, Christian Medical College, Vellore, India^1 | |
| 关键词: Absorption co-efficient; Distilled water; Optical scanners; Peak absorption; Projection image; Quality assurance tools; Radiotherapy applications; Tissue equivalence; | |
| Others : https://iopscience.iop.org/article/10.1088/1742-6596/444/1/012027/pdf DOI : 10.1088/1742-6596/444/1/012027 |
|
| 来源: IOP | |
PDF
|
|
【 摘 要 】
The aim of this study was to investigate the feasibility of using Leuco Crystal Violet (LCV) based micelle gel dosimeter as a quality assurance tool in radiotherapy applications. Basic properties such as absorption coefficient and diffusion of LCV gel phantom over time were evaluated. The gel formulation consisted of 25 mM Trichloroacetic acid, 1mM LCV, 4 mM Triton X-100, 4% gelatin by mass and distilled water. The advantages of using this gel are its tissue equivalence, easy and less preparation time, lower diffusion rate and it can be read with an optical scanner. We were able to reproduce some of the results of Babic et al. The peak absorption was found to be at 600 nm and hence a matrix of yellow LEDs was used as light source. The profiles obtained from projection images confirmed the diffusion of LCV gel after 6 hours of irradiation. Hence the LCV gel phantom should be read before 6 hours post irradiation to get accurate dose information as suggested previously.