Investigation of 3D dosimetry for an anthropomorphic spine phantom | |
Grant, R.^1,2 ; Ibbott, G.^1 ; Yang, J.^1 ; Adamovics, J.^3 ; Followill, D.^1 | |
Department of Radiation Physics, University of Texas MD Anderson Cancer Center, 1515 Holcombe Blvd., Houston, TX 77030, United States^1 | |
University of Texas at Houston, Graduate School of Biomedical Sciences, Houston, TX 77030, United States^2 | |
Heuris Pharma, LLC, 412 Sunset Rd., Skillman, NJ 08558, United States^3 | |
关键词: 3d dosimeters; Clinical environments; Dose distributions; Dose-volume histograms; Radiochromic film; | |
Others : https://iopscience.iop.org/article/10.1088/1742-6596/444/1/012020/pdf DOI : 10.1088/1742-6596/444/1/012020 |
|
来源: IOP | |
![]() |
【 摘 要 】
A new dosimetry insert for the Radiological Physics Center's spine phantom was designed to hold a specially molded dosimeter. The phantom was irradiated with the traditional insert loaded with radiochromic film and TLD, and then with the new 3D dosimetry insert. A comparison with the calculated dose distribution showed that PRESAGE® dosimeter, as well as the film and TLD system, agreed to within ±2mm. Further analysis of the 3D dosimeter, including a measured dose volume histogram, demonstrated the advantages of 3D dosimetry in a clinical environment.