| Korreman, S.S.^1,2,3 | |
| Department of Science, Systems and Models, Roskilde University, Denmark^1 | |
| Department of Radiation Oncology, Rigshospitalet, University of Copenhagen, Denmark^2 | |
| Niels Bohr Institute, University of Copenhagen, Denmark^3 | |
| 关键词: Clinical practices; Measurement methods; Three categories; True 3d; | |
| Others : https://iopscience.iop.org/article/10.1088/1742-6596/444/1/012007/pdf DOI : 10.1088/1742-6596/444/1/012007 |
|
| 来源: IOP | |
PDF
|
|
【 摘 要 】
In this review, 3D dosimetry is divided in three categories; "true" 3D, semi-3D and virtual 3D. Virtual 3D involves the use of measurement arrays either before or after beam entry in the patient/phantom, whereas semi-3D involves use of measurement arrays in phantoms mimicking the patient. True 3D involves the measurement of dose in a volume mimicking the patient.There are different advantages and limitations of all three categories and of systems within these categories. Choice of measurement method in a given case depends on the aim of the measurement, and examples are given of verification measurements with various aims.