Qualification of a vertical cryostat for MLI performances tests between 20 K and 60 K to 4.2 K
Venturi, V.^1^2 ; Koettig, T.^2 ; Chorowski, M.^1 ; Polinski, J.^1 ; Parma, V.^2 ; Kopczynski, A.^1
WUST, Wroclaw University of Science and Technology, Wroclaw, Poland^1
CERN, Organisation Europeen Pour la Recherche Nucleaire, Geneva, Switzerland^2
关键词: Calorimetric measurements;    Controlled temperature;    Cylindrical liquids;    Electrical heater;    Lower temperatures;    Measurements of;    Multi-layer insulation;    Performance measurements;   
Others  :  https://iopscience.iop.org/article/10.1088/1757-899X/502/1/012077/pdf
DOI  :  10.1088/1757-899X/502/1/012077
来源: IOP
PDF
【 摘 要 】

Boil-off calorimetric measurements of Multilayer Insulation (MLI) performance from 20 K-60 K to 4.2 K are being performed at CERN, extending the range of measurements of MLI to cover cryostat applications where thermal shields are operated at lower temperatures than that of LN2. Possible applications include future large accelerators like the Future Circular Collider (FCC). The tests are carried out in an existing vertical test cryostat shielded with a liquid nitrogen cooled vessel and with an inner cylindrical liquid helium tank carrying the wrapped MLI sample. The cryostat has been modified to insert an additional intermediate thermal shield maintained at controlled temperatures in the 20 K-60 K range, by means of regulated power input from an electrical heater. This paper describes the test cryostat and the MLI performance measurement methodology as well as experimental results of the commissioning and calibration of the instrument.Published under licence by IOP Publishing Ltd.

【 预 览 】
附件列表
Files Size Format View
Qualification of a vertical cryostat for MLI performances tests between 20 K and 60 K to 4.2 K 946KB PDF download
  文献评价指标  
  下载次数:13次 浏览次数:8次